Klein-tools CL2300 Manual de usuario Pagina 4

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 24
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 3
600V
400V
CATIII
600V
400V
CATIII
PRESS
PRESS
2300
7
5
2
3
6
1
9
1. AC/DC Current (large):
<
600A
Features:
2. AC / DC Current (small):
<
2000µA
Features:
MAX/MINRANGEHOLD
H. I. Battery Replacement
When
indicator is displayed on the LCD, batteries must be
replaced.
Remove the back screw and replace 2 x AAA batteries.
J. Magnetic Back
Attach instrument to metal for hands-free use.
K. Bar Graph
The bar graph shows an approximate analog representation
of a measurement.
The bar graph responds much faster than the digital display.
The scale of the bar graph is zero to the maximum reading of
the selected range.
L. Temperature Adjustment
Remove battery cover.
Place temperature probe in distilled
water with stirred crushed ice
(a standard for 0°C and 32°F).
Use a fi ne tip screwdriver to adjust
the screw recessed in the bottom
right corner of the battery cavity as
shown until the display reads 32°F.
FUNCTION INSTRUCTIONS
Select AC or DC current source.
Current above 2000µA will damage instrument.
MAX/MINRANGEHOLD
Center wire in guides for best accuracy.
Opposing currents cancel
(use line-splitter when necessary)
.
ZERO (DC)
Vista de pagina 3